Measuring range (Auto range) |
100.0M??50V? 200.0M??100V? 250.0M??125V? 500.0M??250V? 20.00G??500V? 40.00G??1000V? |
---|---|
Mid-scale value | 2M??50V? 5M??100V/125V? 10M??250V? 100M??500V? 200M??1000V? |
First effective measuring range | 0.100 ? 10.00M??50V? 0.100 ? 20.00M??100V? 0.100 ? 25.00M??125V? 0.100 ? 50.0M??250V? 0.100 ? 500M??500V? 0.100 ? 1000M??1000V? |
Accuracy | ±2%rdg ±2dgt |
Second effective measuring ranges | 0.050 ~ 0.099M??50V/100V/125V/250V/500V/1000V? 10.01 ~ 100.0M??50V? 20.01 ~ 200.0M??100V? 25.01 ~ 250.0M??125V? 50.1 ~ 500M??250V? 501 ~ 2000M??500V? 1001 ~ 4000M??1000V? |
Accuracy | ±5%rdg (0.050 ~ 0.099M?: ±2%rdg±4dgt) |
Rated current | 1.0 ~ 1.1mA |
Output short circuit current | 1.5mA max |
Auto range | 40.00/400.0/4000? |
---|---|
Accuracy | ±2.5%rdg ±8dgt |
Output voltage on open circuit | 5V?4~6.9V? |
Output short circuit current | 200mA max |
Voltage range | AC 2.0~600V?45~65Hz?? DC -2.0~-600V +2.0~+600V |
---|---|
Accuracy | ±1%rdg ±4dgt |
Applicable standards | IEC61010 CAT? 600V/CAT IV 300V IEC61557-1,2,4 IEC61326-1,-2-2 IEC60529?IP40? |
---|---|
Power source | LR6/R6?AA??1.5V?x4 |
Communication interface | USB , Bluetooth®4.0LE(Bluetooth®SMART) |
Dimensions / Weight | 97(L) x 156(W) x 46(D) mm / 490g approx?including battery? |
Included Accessories | 7260?Test leads with remote control switch? 7261A?Test leads with alligator clip? 8017A?Extension prod long? 9173?Carrying case? 9121?Shoulder strap? LR6?AA??1.5V?x4 Instruction manual |
Optional Accessories | 7243A?L-shaped probe? 8016?Hook type prod? 8212-USB?USB adaptor with "KEW Report (software)" ? |