High Accuracy, High Performance and Reliable Measurements
Safety design for industrial use
Reliable support for data management
Detection mode | MEAN/RMS (switch) |
---|---|
DC V | 600.0mV/6.000/60.00/600.0/1000V (Input impedance: 10M? [600mV/60/600/1000V], 11M? [6V]) ±0.09%rdg±2dgt (Basic accuracy) |
AC V [RMS] | 600.0mV/6.000/60.00/600.0/1000V (Input impedance: 10M?<200pF [600mV], 11M?<50pF [6V], 10M?<50pF [60/600/1000V]) ±0.5%rdg±5dgt (Basic accuracy) |
AC V [MEAN] | 600.0mV/6.000/60.00/600.0/1000V (Input impedance: 10M?<200pF [600mV], 11M?<50pF [6V], 10M?<50pF [60/600/1000V]) ±0.5%rdg±5dgt (Basic accuracy) |
DC A | 600.0/6000µA/60.00/440.0mA/6.000/10.00A ±0.2%rdg±2dgt (Basic accuracy) |
AC A [RMS] | 600.0/6000µA/60.00/440.0mA/6.000/10.00A ±0.75%rdg±5dgt (Basic accuracy) |
? | 600.0?/6.000/60.00/600.0k?/6.000/60.00M? ±0.4%rdg±1dgt (Basic accuracy) |
Continuity buzzer | 600.0? (The buzzer turns on for resistances lower than 50±30?) |
Diode test | 2.000V ±1%rdg±2dgt Open curcuit voltage: <3.5V (Approx. 0.5mA Measuring Current) |
Capacitance | 10.00/100.0nF/1.000/10.00/100.0/1000µF ±2%rdg±5dgt (Basic accuracy) |
Frequency | 10.00 - 99.99/90.0 - 999.9Hz/0.900 - 9.999/9.00 - 99.99kHz ±0.02%rdg±1dgt (Basic accuracy) |
Temperature | -50 - 600ºC ±2%rdg±2ºC (with the use of K-type Temperature probe) |
Other functions | Data Hold (D/H), Auto Hold (A/H), Range Hold (R/H), Maximum value (MAX) *, Minimum value (MIN) *, Average value (AVG) *, Zero Adjustment (Capacitor, Resistance), Save to Memory *, LCD backlight |
Applicable standards | IEC 61010-1 CAT IV 600V, CAT III 1000V Pollution degree 2, IEC 61010-031, IEC 61326-1 (EMC) |
Power source | R6 (1.5V)×4 (Auto power off: approx. 20 minutes) |
Dimensions | 192(L)×90(W)×49(D) mm |
Weight | Approx. 560g (including batteries) |
Included Accessories | 7220A (Test leads), R6×4, Instruction manual, 8926 (Fuse [440mA/1000V])×1 (included) 8927 (Fuse [10A/1000V])×1 (included) |